Author Interviews, Blood Pressure - Hypertension, JAMA, Pediatrics / 08.08.2023

MedicalResearch.com Interview with: Tammy M. Brady, MD, PhD  (she/her/hers) Vice Chair for Clinical Research, Dept of Pediatrics Associate Director, Welch Center for Prevention, Epidemiology, and Clinical Research Associate Professor of Pediatrics, Division of Pediatric Nephrology Medical Director, Pediatric Hypertension Program Johns Hopkins University Baltimore, MD 21287     MedicalResearch.com: What is the background for this study? What are the main findings? Response: Accurate BP measurement is key to identification and treatment of hypertension which serves ultimately to prevent cardiovascular disease.  Our study describes substantial measurement error that can occur in a common office and home BP measurement scenario: use of a regular cuff size for all individuals regardless of arm size.  Many office triage measurements occur without individualized cuff selection and most home BP devices come with one cuff size – and our study shows that using a regular cuff size for people who have larger arms – those who require a large adult cuff or an extra-large adult cuff – can lead to blood pressure readings that are almost 5 and 20 mmHg greater than their actual BP, respectively.  Those require a small adult cuff can have BP readings that are almost 4 mmHg lower than their actual BP. (more…)